FUSION OMNI-TOME
Report
- Report Number
- 1037905-2011-00350
- Event Type
- Malfunction
- Date Received
- May 27, 2011
- Date of Event
- April 28, 2011
- Report Date
- April 28, 2011
- Manufacturer
- WILSON-COOK MEDICAL, INC.
- Product Code
- KNS
- PMA / PMN Number
- K052051
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
THE INFO IN THIS REPORT WAS PROVIDED TO US BY THE COOK FACILITY IN (B)(6) ON BEHALF OF A DISTRIBUTOR IN (B)(4). (B)(6). EVAL: A PRODUCT EVAL WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVAL. THE REPORT COULD NOT BE CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD AND SAMPLE TEST FROM THIS LOT COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. AFTER A REVIEW OF THE ACCOUNT ORDERING AND SHIPPING HISTORY, THE LOT NUMBER INVOLVED COULD NOT BE DETERMINED. CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE, THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVAL. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. CORRECTIVE ACTION: NO CORRECTIVE ACTION WARRANTED AT THIS TIME BECAUSE THIS REPORT COULD NOT BE VERIFIED. HOWEVER, THIS TYPE OF OCCURRENCE HAS BEEN BROUGHT TO THE ATTENTION OF THE APPROPRIATE INTERNAL PERSONNEL IN AN EFFORT TO HEIGHTEN AWARENESS. NO FURTHER ACTION WAS TAKEN AT THIS TIME. CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION OMNI-TOME. THE SPHINCTEROTOME WAS INTRODUCED INTO THE ACCESSORY CHANNEL OF THE ENDOSCOPE. WHEN THE SPHINCTEROTOME EXITED THE ENDOSCOPE THE CUTTING WIRE ORIENTATION WAS INCORRECT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADD'L PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FUSION OMNI-TOME | KNS, ENDOSCOPIC ELECTROSURGICAL UNIT AND ACCESSORIES | KNS | WILSON-COOK MEDICAL, INC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ENDOSCOPE (UNK TYPE)| ELECTROSURGICAL UNIT (UNK TYPE)| ACTIVE CORD (UNK TYPE) |