FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 2112481
·
Received May 24, 2011
Report
- Report Number
- 3015876-2011-00411
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- April 27, 2011
- Report Date
- April 27, 2011
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K991910
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE SYSTEM/MEMORY PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED SYSTEM/MEMORY PCB ASSEMBLY AND DETERMINED THE CAUSE OF THE MALFUNCTION TO BE A SHORTED CAPACITOR, DESIGNATOR C318.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES | MKJ | PHYSIO-CONTROL, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |