FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 2112473 · Received April 28, 2011

Report

Report Number
9615739-2011-00107
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
April 5, 2011
Report Date
April 5, 2011
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE HEAD DRIVE WAS DAMAGED BY A STUCK HEAD UP BUTTON IN THE LEFT SIDERAIL. HE REPLACED THE HEAD DRIVE MOTOR AND THE CAREGIVER CONTROLS ON THE LEFT SIDERAIL TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT STATED THE HEAD SECTION IS FULLY RAISED AND WILL NOT LOWER ELECTRICALLY OR BY ACTIVATING THE CPR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1