FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1,10,MTX,MG

MDR report key: 21124618 · Received January 10, 2025

Report

Report Number
0002023141-2025-00091
Event Type
Injury
Date Received
January 10, 2025
Date of Event
November 28, 2024
Report Date
May 15, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019829
PMA / PMN Number
K101977
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PERI-IMPLANTITIS IS AN INFECTIOUS DISEASE THAT CAUSES INFLAMMATION OF THE GUM AND THE BONE STRUCTURE AROUND A DENTAL IMPLANT. CHRONIC INFLAMMATION CAUSES BONE LOSS, WHICH CAN LEAD TO IMPLANT FAILURE THUS ITS REMOVAL. INVESTIGATIONS PERFORMED FOR HUNDREDS OF EVENTS WHERE EITHER OF THESE CONDITIONS, INFECTION, OR BONE LOSS, OR BOTH, HAVE BEEN REPORTED, HAVE CONCLUDED THAT THE LIKELY CAUSE(S) FOR THE IMPLANT FAILURE IN RELATION TO THESE CONDITIONS ARE EXTERNAL FACTORS. THOSE INCLUDE MEDICAL CONDITIONS (E.G., DIABETES, BRUXISM, ETC.), PATIENT HABITS (E.G., SMOKING, BAD ORAL HYGIENE ROUTINE) AND USER ERROR (E.G., SURGICAL TECHNIQUE). THERE HAS NOT BEEN ANY INSTANCE WHERE EITHER INFECTION AND/OR BONE LOSS, AND WHEN RIGHT CONDITIONS PREVAIL AND LED TO PERI-IMPLANTITIS WHETHER REPORTED OR NOT ALONG WITH THE OTHER TWO, HAVE RESULTED FROM A MANUFACTURING OR DESIGN DEFECT. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO INFECTION AND RESULT IN BONE LOSS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. THE ANALYSIS AND RESULT OF INVESTIGATIONS AND THE ANALYSIS OF PROBABILITY PREVIOUSLY DESCRIBED ARE CONTAINED IN THE SUMMARY INVESTIGATION REPORTS PERFORMED FOR BONE LOSS AND INFECTION, WHICH ARE ATTACHED. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/ OR ANY FURTHER ESCALATIONS. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1242334. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD (OP160) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1242334 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿PERI-IMPLANTITIS.¿ AS DOCUMENTED IN THE SUMMARY INVESTIGATIONS, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATIONS, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS THEY ARE A MEDICAL CONDITION. MONTHLY TRENDING TO DATE HAS NOT IDENTIFIED ANY STATISTICAL TRIGGERS SUGGESTING POTENTIAL DESIGN OR MANUFACTURING RELATED ISSUES. PREVIOUSLY COMPLETED INVESTIGATIONS FOR INFECTION, BONE LOSS, AND/OR PERI-IMPLANTITIS PROBLEMS REPORTED IN ASSOCIATION WITH IMPLANT FAILURE HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES IMPACTING THE MANUFACTURING AND STERILIZATION PROCESSES. THEREFORE, ESCALATION TO CAPA IS NOT REQUIRED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED.

Additional Manufacturer Narrative · 0

THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. D4: LOT/SERIAL #. D4: DEVICE EXPIRATION DATE. D4: DEVICE UDI NUMBER. D9: DEVICE AVAILABILITY. G3: DATE RECEIVED BY MANUFACTURER. G6: CHECKED "FOLLOW-UP". H2: CHECKED FOLLOW-UP TYPE. H4: DEVICE MANUFACTURER DATE. H6: ENTERED EVALUATION CODES. H10: ADDED MANUFACTURER NARRATIVE.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT IMPLANT WAS REMOVED DUE TO PERI-IMPLANTITIS, BONE LOSS. THE PATIENT WAS RECEIVED 1 IMPLANT PLACEMENT, ALREADY HAD THE FINAL RESTORATION. 1 YEAR LATER, THE PATIENT CAME BACK FOR THE ANNUAL CHECK-UP. THE BONE LOSS (7MM) AND PERI-IMPLANTITIS WAS RECOGNIZED. AFTER THE DISCUSSION, THE PATIENT AGREED TO REMOVE THE IMPLANT. TOOTH SITE # 24.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1375707 IMP,TSV,4.1,10,MTX,MG DENTAL IMPLANT DZE ZIMMER DENTAL 1242334 00889024019829

Patients

Seq Age Sex Outcome Treatment
1 52 YR Male Required Intervention