FDA Adverse Event Malfunction Summary report: N

CENTURY BED

MDR report key: 2112425 · Received April 28, 2011

Report

Report Number
1824206-2011-02435
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
April 4, 2011
Report Date
April 4, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BED CABLE PINCHED BETWEEN THE KNEE SUPPORT AND THE INTERMEDIATE FRAME FOR KNEE UP CAUSING THE MOTOR TO CONTINUOUSLY RUN. THE TECHNICIAN REPLACED THE CABLE TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE KNEE FUNCTION RUNS UP WITH NO BUTTONS ENGAGED. HE UNPLUGGED BOTH SIDERAIL AND REPLACED THE BOARD BUT THE KNEE STILL RAN UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTURY BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 840

Patients

Seq Age Sex Outcome Treatment
1