FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2112380 · Received May 26, 2011

Report

Report Number
2032227-2011-01334
Event Type
Injury
Date Received
May 26, 2011
Date of Event
May 17, 2011
Report Date
May 18, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE LEVELS OVER 500 MG/DL. THE CUSTOMER'S LAST INFUSION SET CHANGE WAS ON (B)(6), 2011. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT FAILED THE HIGH PRESSURE TEST. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-722LNAS

Patients

Seq Age Sex Outcome Treatment
1 54 YR Hospitalization