DEXTRUS 4137
Report
- Report Number
- 1028232-2011-01190
- Event Type
- Injury
- Date Received
- May 27, 2011
- Date of Event
- April 5, 2011
- Report Date
- May 24, 2011
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR (RV) LEAD DEVELOPED A HIGH THRESHOLD MEASUREMENT OF 2.8 VOLTS AND INTERMITTENT LOSS OF CAPTURE POST IMPLANT. A LEAD REVISION PROCEDURE WAS SCHEDULED FOR THE FOLLOWING DAY. IT WAS NOTED THAT THE RV LEAD HAD PERFORATED AND THE PATIENT WAS TAKEN TO THE OPERATING ROOM FOR AN EMERGENCY PERICARDIAL WINDOW DUE TO A LARGE PERICARDIAL EFFUSION. THE BLOOD WAS REMOVED FROM THE PERICARDIAL SAC AND AN ATTEMPT TO REPOSITION THE LEAD WAS UNSUCCESSFUL. THE RV LEAD WAS TAKEN OUT OF THE POCKET AND VISUAL INSPECTION NOTED AN ISSUE WITH THE HELIX MECHANISM. THE BOSTON SCIENTIFIC SALES REPRESENTATIVE STATED THAT THE INTERMITTENT LOSS OF CAPTURE AND HELIX ISSUE WERE LIKELY DUE TO OVER TURNING OF THE MECHANISM BY THE PHYSICIAN WHEN ATTEMPTING TO REPOSITION. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4137 | PACER LEAD | NVZ | BIOTRONIK SE & CO. KG | 358764 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |