FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2112121 · Received June 2, 2011

Report

Report Number
6000001-2011-07022
Event Type
Malfunction
Date Received
June 2, 2011
Date of Event
May 6, 2011
Report Date
May 6, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION AND THE REPORTED CONDITION WAS CONFIRMED IN THE EVENT HISTORY BUT NOT DUPLICATED. THIS COMPLAINT IS AN ANCILLARY OF (B)(4). THE DEVICE PASSED TESTING AND NO FAILURE COULD BE DETECTED. THE CAUSE OF THE EVENT IS UNKNOWN. NO REPAIRS WERE PERFORMED TO CORRECT THE REPORTED CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND WITH 3 OCCURRENCES OF AN AIR IN LINE SET, WHICH CONSTITUTE POSSIBLE FALSE AIR IN LINE ALARMS. THERE WAS NO PATIENT INVOLVEMENT, INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS INVOLVED AN UNREMEDIATED INFUSION PUMP WITH USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.04.00.