HOMECHOICE
Report
- Report Number
- 1423500-2011-07027
- Event Type
- Malfunction
- Date Received
- June 2, 2011
- Date of Event
- April 9, 2011
- Report Date
- May 11, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE CAUSE OF THE IIPV IDENTIFIED IN THE DEVICE LOG WAS UNDETERMINED. A LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED THE DEVICE FAILED PNEUMATIC TESTING. A RELATIONSHIP BETWEEN IIPV AND THE PNEUMATIC FAILURE IS UNDETERMINED. HOWEVER, THE DEVICE WAS REWORKED AND PASSED ALL SUBSEQUENT TESTING. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH ((B)(4)).
(B)(4).THE EVALUATION OF THE DEVICE HAS BEGUN; HOWEVER, HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN PATIENT THERAPY LOG ON (B)(6) 2011 STARTED AT 11:06 WITH DRAIN VOLUME OF 4986 ML DURING INITIAL DRAIN. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED. THIS EVENT MEETS OVERFILL CRITERIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |