FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2112100 · Received June 2, 2011

Report

Report Number
1423500-2011-07026
Event Type
Malfunction
Date Received
June 2, 2011
Date of Event
April 14, 2011
Report Date
May 11, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE EVALUATION OF THE DEVICE HAS BEGUN; HOWEVER, HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE CAUSE OF THE IIPV IDENTIFIED IN THE DEVICE LOG WAS UNDETERMINED. A LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED THE DEVICE FAILED PNEUMATIC TESTING. A RELATIONSHIP BETWEEN IIPV AND THE PNEUMATIC FAILURE IS UNDETERMINED. HOWEVER, THE DEVICE WAS REWORKED AND PASSED ALL SUBSEQUENT TESTING. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (RTB-CAPA-(B)(4)).

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN PATIENT THERAPY LOG ON (B)(6) 2011 STARTED AT 12:44:08 WITH DRAIN VOLUME OF 5173 ML DURING INITIAL DRAIN. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED. THIS EVENT MEETS OVERFILL CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 45 YR