DAVINCI XI
Report
- Report Number
- 2955842-2025-00274
- Event Type
- Malfunction
- Date Received
- January 9, 2025
- Date of Event
- December 16, 2024
- Report Date
- December 16, 2024
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110744
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- 003
Narratives
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE RESEATED THE CONNECTOR ON THE HIGH-RESOLUTION STEREO VIEWER (HRSV). THE HRSV DISPLAYED NORMALLY. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE THAT THE SURGEON SIDE CONSOLE'S (SSC) HIGH RESOLUTION STEREO VIEWER (HRSV) HAD NO IMAGE IN THE RIGHT EYE. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) RECOMMENDED HARD POWER CYCLING THE SYSTEM AND RESEATING THE BLUE FIBER CABLE, BUT THE ISSUE REMAINED. THE SITE CONVERTED TO ANOTHER DA VINCI SYSTEM. THERE WERE NO REPORTS OF PATIENT INJURY. ISI RECEIVED THE FOLLOWING ADDITIONAL INFORMATION VIA FOLLOW-UP: THE SYSTEM DID NOT PRESENT ANY FAULTS WHEN INITIALLY POWERED ON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 364753 | DAVINCI XI | SURGEON SIDE CONSOLE, SMART PEDALS | NAY | INTUITIVE SURGICAL, INC | 380677-17 | N/A | 00886874110744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |