FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 21120501 · Received January 9, 2025

Report

Report Number
2955842-2025-00274
Event Type
Malfunction
Date Received
January 9, 2025
Date of Event
December 16, 2024
Report Date
December 16, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110744
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE RESEATED THE CONNECTOR ON THE HIGH-RESOLUTION STEREO VIEWER (HRSV). THE HRSV DISPLAYED NORMALLY. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE THAT THE SURGEON SIDE CONSOLE'S (SSC) HIGH RESOLUTION STEREO VIEWER (HRSV) HAD NO IMAGE IN THE RIGHT EYE. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) RECOMMENDED HARD POWER CYCLING THE SYSTEM AND RESEATING THE BLUE FIBER CABLE, BUT THE ISSUE REMAINED. THE SITE CONVERTED TO ANOTHER DA VINCI SYSTEM. THERE WERE NO REPORTS OF PATIENT INJURY. ISI RECEIVED THE FOLLOWING ADDITIONAL INFORMATION VIA FOLLOW-UP: THE SYSTEM DID NOT PRESENT ANY FAULTS WHEN INITIALLY POWERED ON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
364753 DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS NAY INTUITIVE SURGICAL, INC 380677-17 N/A 00886874110744

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES