ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-01689
- Event Type
- Malfunction
- Date Received
- June 2, 2011
- Date of Event
- April 28, 2011
- Report Date
- May 4, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K922823/A007
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THE PATIENT SAMPLES WERE PERFORMED UTILIZING CALIBRATOR REQUIRING FREEZER STORAGE; HOWEVER, THE CALIBRATOR STORAGE INFORMATION WAS NOT PROVIDED. A SYSTEM CHECK PERFORMED ON (B)(6) 2011 WAS WITHIN ESTABLISHED SPECIFICATIONS. CALIBRATION DATED (B)(6) 2011 PASSED. QC WAS WITHIN ESTABLISHED RANGES. SERVICE INFORMATION WAS NOT SUPPLIED. A DEFINITIVE ROOT CAUSE IS UNKNOWN. ADDITIONAL MDRS FILED FOR THIS EVENT TO ADDRESS EACH DATE THAT THE RESULTS IN QUESTION WERE GENERATED: 2122870-2011-01684, 2122870-2011-01686, 2122870-2011-01687, 2122870-2011-01688, 2122870-2011-01690, 2122870-2011-01691.
A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI), REPORTING THAT THEY BELIEVE RESULTS FOR FREE T4 (FT4) ASSAY PERFORMED ON ACCESS 2 IMMUNOASSAY SYSTEM RECOVERED GREATER THAN THE BCI REFERENCE RANGE FOR MULTIPLE PATIENT SAMPLES TESTED BETWEEN (B)(6) 2011 THROUGH (B)(6) 2011. SPECIFIC DATES AND TIME OF EACH SAMPLE WERE NOT PROVIDED, ONLY RANGE OF DATES. PATIENT RESULTS ARE PROVIDED. THESE RESULTS CORRESPOND TO TSH RESULTS WITHIN BCI'S NORMAL REFERENCE RANGE. THE FT4 RESULTS WERE REPORTED OUT OF THE LAB. THERE WERE NO REPORTS REGARDING AN EFFECT TO A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS® 2 IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | ACCESS® 2 IMMUNOASSAY SYSTEM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |