FDA Adverse Event Injury Summary report: N

GUARDIAN EZ-CARE

MDR report key: 2111828 · Received November 4, 2008

Report

Report Number
3006459587-2008-00075
Event Type
Injury
Date Received
November 4, 2008
Date of Event
May 21, 2008
Report Date
November 3, 2008
Manufacturer
PROFIT COUNTRY ENTERPRISES LTD.
Product Code
INM
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ASSESSMENT IS THAT THE SUPPORTING BACK CLAMP BROKE AWAY FROM THE CHAIR. THE BREAK WAS NOT THE RESULT OF DESIGN OR FAULTY ASSEMBLY. IT APPEARS THE USER WAS USING THE BACK OF THE CHAIR AS A SUPPORT WHEN EXITING THE SHOWER/BATH AREA. THE CHAIR IS DESIGNED TO SUPPORT A PTS DISTRIBUTED WEIGHT. USING THE BACK OF THE CHAIR AS A SUPPORT CAUSED AN OVERLOAD TO THE BACK AND THE SCREWS POPPED AWAY FROM THE ALUMINUM FRAME.

Description of Event or Problem · 1

PT FELL (NO INJURY) WHEN BACK REST SCREWS PULLED AWAY FROM THE SHOWER CHAIR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GUARDIAN EZ-CARE SHOWER CHAIR WITH BACK INM PROFIT COUNTRY ENTERPRISES LTD. 30400-04

Patients

Seq Age Sex Outcome Treatment
1 UNK