FDA Adverse Event Injury Summary report: N

GORE EXCLUDER AAA ENDOPROSTHESIS

MDR report key: 2111771 · Received May 25, 2011

Report

Report Number
2953161-2011-00122
Event Type
Injury
Date Received
May 25, 2011
Date of Event
May 2, 2011
Report Date
May 24, 2011
Manufacturer
W. L. GORE & ASSOCIATES
Product Code
MIH
PMA / PMN Number
P020004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE MFG PAPERWORK HAS BEEN CONDUCTED. THE REVIEW OF THE MFG PAPERWORK VERIFIED THAT THIS LOT MET ALL PRE-RELEASE SPECS. ANEURYSM GROWTH IN THE ABSENCE OF ENDOLEAK HAS BEEN DEFINED AS ENDOTENSION. ONE HYPOTHESIS FOR THE SOURCE OF ENDOTENSION IS THE TRANSMURAL MOVEMENT OF SEROUS FLUID ACROSS THE MATERIAL USED TO FABRICATE DEVICES USED TO TREAT THE AORTIC ABDOMINAL ANEURYSM. IN RESPONSE TO THIS ISSUE, GORE ELECTED TO PROVIDE A DESIGN ENHANCEMENT TO THE ORIGINAL GORE EXCLUDER BIFURCATED ENDOPROSTHESIS.

Description of Event or Problem · 1

ON (B)(6) 2003, THE PT WAS IMPLANTED WITH TWO GORE EXCLUDER BIFURCATED ENDOPROSTHESIS TO TREAT AN ABDOMINAL AORTIC ANEURYSM. ON (B)(6) 2011, CT SHOWED ANEURYSM GROWTH OF APPROX 5 MM WITH NO EVIDENT ENDOLEAK AS COMPARED WITH CT OF (B)(6) 2009. THE PHYSICIAN IS PRESENTLY MONITORING THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GORE EXCLUDER AAA ENDOPROSTHESIS MIH/SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W. L. GORE & ASSOCIATES WLG425 022734205

Patients

Seq Age Sex Outcome Treatment
1 70 YR Other (B)(4)