FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM

MDR report key: 2111739 · Received June 2, 2011

Report

Report Number
2050012-2011-02112
Event Type
Malfunction
Date Received
June 2, 2011
Date of Event
April 12, 2011
Report Date
April 12, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PER CUSTOMER, THEY HAVE HAD TG SUPPRESSION ISSUES IN THE PAST, AND THEIR FIELD SERVICE ENGINEER (FSE) HAS RECENTLY REPLACED THE ENTIRE WASH TOWER. BCI CUSTOMER TECHNICAL SUPPORT (CTS) HELPED THE CUSTOMER TROUBLESHOOT THE HARDWARE (CLEAN THE CARTRIDGE CHEMISTRY SAMPLE PROBE AND COLLAR WASHER, REPLACE THE REAGENT MIXER PADDLE AND THE SYRINGE PLUNGER RODS). CTS FOUND THE CUSTOMER HAD PUSHED THE WIPER UP TOO FAR UP AGAINST THE PHYSICAL STOP ON THE PROBE. THE CUSTOMER WAS INFORMED THAT THE WIPER SHOULD BE FLUSH WITH THE TIP OF THE PROBE. FIELD SERVICE ENGINEER (FSE) VISITED THE SITE 3 TIMES, PERFORMED TROUBLESHOOTING, REPLACED HARDWARE AS NECESSARY AND PERFORMED SEVERAL REPAIRS. ON THE THIRD VISIT, (B)(4)-2011 FSE REPLACED WASH VACUUM PROBES. REPLACED REAGENT PROBE A WASH COLLAR AND PERFORMED ALIGNMENTS. RAN ALL PVTS WITH NO FAILURES, RAN REAGENT PROBE CARRYOVER 4 TIMES WITH NO FAILURES. THE REPAIRS PERFORMED BY THE FSE RESOLVED THE ISSUE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REPORTING HIGH RESULTS FOR TRIGLYCERIDES (TG) GENERATED BY THE UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEM. THE FALSE HIGH RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY. TG WAS 297MG/DL AND UPON RERUN ON OTHER ANALYZER WAS 124MG/DL. THE CUSTOMER DID NOT REPORT ANY DEATH, INJURY OR CHANGE TO PATIENT TREATMENT ATTRIBUTED TO OR CONNECTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1