FDA Adverse Event Malfunction Summary report: N

PERMANENT CAUTERY HOOK INSTRUMENT

MDR report key: 2111686 · Received June 2, 2011

Report

Report Number
2955842-2011-00164
Event Type
Malfunction
Date Received
June 2, 2011
Date of Event
April 13, 2011
Report Date
May 3, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K002489
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THE HOOK BROKEN OFF OF THE INSTRUMENT. THE HOOK HAD FRACTURED WITHIN THE YAW PULLEY, AT THE CROSS SECTION OF THE HOLE FEATURE. THE DETACHED HOOK WAS NOT RETURNED. ENGINEERING ALSO OBSERVED THAT ONE SIDE OF THE CERAMIC SLEEVE IS BROKEN OFF AND MISSING. THE SIZE OF THE MISSING PIECE IS APPROXIMATELY .155 X .075. IT WAS DETERMINED THAT THE DAMAGE TO THE HOOK AND SLEEVE MAY BE DUE TO OVERLOADING AT THE TIP OR OTHER MISHANDLING/MISUSE. NO OTHER DAMAGE FOUND.

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED OR ADDITIONAL INFORMATION IS RECEIVED. PER THE INFORMATION PROVIDED BY THE INITIAL REPORTER, THE BROKEN INSTRUMENT PIECE WAS SUCCESSFULLY RETRIEVED FROM THE PATIENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI SIGMOID COLECTOMY THE TIP OF THE PERMANENT CAUTERY HOOK INSTRUMENT BROKE AND A PIECE FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH A REPLACEMENT INSTRUMENT OF THE SAME TYPE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERMANENT CAUTERY HOOK INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420183-05 M10100505 402