COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-06982
- Event Type
- Malfunction
- Date Received
- June 2, 2011
- Date of Event
- January 14, 2011
- Report Date
- May 4, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE DEPLETED MAIN BATTERIES. TO CORRECT THE CONDITION, THE MAIN BATTERIES WERE REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ONSITE AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
DURING REVIEW OF THE PUMP'S EVENT HISTORY, BAXTER (B)(4) DISCOVERED A COLLEAGUE INFUSION PUMP EXPERIENCED A BATTERY DEPLETED ALARM, WHICH INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.08.92.