FDA Adverse Event Injury Summary report: N

FLEXICAP

MDR report key: 2111590 · Received June 2, 2011

Report

Report Number
1423500-2011-06978
Event Type
Injury
Date Received
June 2, 2011
Date of Event
December 1, 2010
Report Date
May 13, 2011
Manufacturer
BAXTER HEALTHCARE - SWINFORD
Product Code
KDJ
PMA / PMN Number
K972579
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE OF THE REPORTED CONDITION OF PERITONITIS WAS USE ERROR- POOR ASEPTIC TECHNIQUE. A BATCH REVIEW WAS PERFORMED FOR THE POTENTIALLY ASSOCIATED LOT NUMBERS (10E17H25, 10G15H25). THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE. A LABELING REVIEW WAS PERFORMED AND THE LABELING WAS FOUND TO PROVIDE AMPLE INSTRUCTIONS RELATED TO PREVENTION OF THE USE ERROR IN ASEPTIC TECHNIQUE. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Additional Manufacturer Narrative · 1

(B)(4). THIS IS REPORT 3 OF 4 INVOLVED IN THIS PERITONITIS EVENT. AS THE DATE OF ONSET OF THIS PERITONITIS EPISODE IS UNKNOWN AND PATIENTS DISCARD SUPPLIES AFTER EACH THERAPY, THE SAMPLE WAS NOT REQUESTED. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THIS REPORT WAS RECEIVED FROM GLOBAL PHARMACOVIGILANCE. THIS IS A SPONTANEOUS REPORT BY A CONSUMER FROM THE USA OF PATIENT MADE MISTAKE/TOUCH CONTAMINATION AND PERITONITIS WITH CULTURE POSITIVE FOR (B)(6) WITH DIANEAL PD4 AMBUFLEX THERAPY. ON AN UNREPORTED DATE, THE PATIENT BEGAN TREATMENT WITH DIANEAL PD4 AMBUFLEX (LOT NUMBER, DOSE AND FREQUENCY NOT REPORTED) INTRAPERITONEALLY (IP) FOR PERITONEAL DIALYSIS (PD). DURING A CALL WITH BAXTER CUSTOMER SERVICES, THE FOLLOWING INFORMATION WAS REPORTED. ON AN UNREPORTED DATE, THE PATIENT MADE A MISTAKE/TOUCH CONTAMINATION. IN (B)(6) 2010, THE PATIENT EXPERIENCED PERITONITIS. THE CAUSE OF THE PERITONITIS WAS PATIENT MADE A MISTAKE/TOUCH CONTAMINATION. THE PATIENT WAS NOT HOSPITALIZED. TREATMENT WAS NOT REPORTED. ON AN UNREPORTED DATE, THE PATIENT HAD RECOVERED FROM THE PERITONITIS AND PD THERAPY WAS ONGOING. THE OUTCOME FOR THE PATIENT MADE MISTAKE/TOUCH CONTAMINATION WAS NOT REPORTED. AN OPINION OF CAUSALITY WAS NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXICAP SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - SWINFORD

Patients

Seq Age Sex Outcome Treatment
1 70 YR Other DIANEAL PD4 AMBUFLEX| HOMECHOICE DEVICE