FDA Adverse Event Malfunction Summary report: N

ARCHITECT C8000 SYSTEM

MDR report key: 2111522 · Received June 2, 2011

Report

Report Number
1628664-2011-00252
Event Type
Malfunction
Date Received
June 2, 2011
Report Date
May 18, 2011
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
PMA / PMN Number
K980367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4), CONCOMITANT MEDICAL PRODUCTS: ARCHITECT ICT PRE-AMP BOARD PHOTO COUPLER LOTS 0834 AND 0840. (B)(4), NEW (B)(4) PROCESS IMPACTED THE LONG TERM RELIABILITY OF THE PHOTO COUPLERS WHICH DEGRADE SLOWLY OVER TIME. THE CAUSE OF THE ARCHITECT ANALYZER PHOTO COUPLER DEGRADATION ISSUE WAS DUE TO A NEW MANUFACTURING (B)(4) PROCESS IMPLEMENTED BY THE SUPPLIER. ABBOTT ISSUED A PRODUCT CORRECTION LETTER TO ALL CUSTOMERS WITH AFFECTED INSTRUMENT SERIAL NUMBERS AND IMPLEMENTED A MANDATORY TECHNICAL SERVICE BULLETIN TO INSPECT ICT PRE AMP BOARDS AND REPLACE IF NEEDED.

Description of Event or Problem · 1

ABBOTT FIELD SERVICE INSPECTED THE ARCHITECT ANALYZER AT THE CUSTOMER FACILITY PER MANDATORY TECHNICAL SERVICE BULLETIN AND REPLACED THE ARCHITECT ICT PRE-AMP BOARD AS REQUIRED. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT C8000 SYSTEM AUTOMATED CHEMISTRY ANALYZER JJE ABBOTT MANUFACTURING, INC. C8000

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT ICT PRE-AMP BOARD PHOTO COUPLER