UNICEL® DXC 600 PRO® SYNCHRON CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-02071
- Event Type
- Malfunction
- Date Received
- June 2, 2011
- Date of Event
- May 4, 2011
- Report Date
- May 4, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
SAMPLE COLLECTION AND CENTRIFUGATION INFORMATION WERE NOT PROVIDED. CONTROLS RUN BEFORE AND AFTER INCIDENT WERE WITHIN ESTABLISHED RANGES. ON (B)(4) 2011, BCI CUSTOMER TECHNICAL SUPPORT (CTS) DISCUSSED PROPER SAMPLE HANDLING WITH CUSTOMER. CTS ADVISED CUSTOMER TO PERFORM CUVETTE WASH PROCEDURE, HAD CUSTOMER PERFORMED QC AND RAN PRECISION TEST. ON (B)(4) 2011, CUSTOMER REPORTED THAT INSTRUMENT WAS GENERATING MULTIPLE REAGENT SYRINGE MOTION ERROR. ON (B)(4) 2011, FIELD SERVICE ENGINEER (FSE) EVALUATED INSTRUMENT AND FOUND LEAK FROM REAGENT PROBE A, AND PUDDLE OF WATER ON THE COVER BENEATH THE PROBE. FSE REPLACED COMPONENTS TO ADDRESS THIS ISSUE, PERFORMED TOP END ALIGNMENT, AND VERIFIED WASH STATION PERFORMANCE, RAN CARRYOVER PROCEDURE TEST AND RESULTS MET SPECIFICATIONS.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT THAT THE UNICEL DXC 600 PRO INSTRUMENT GENERATED TWO (2) FALSE HIGH % HEMOGLOBIN ((B)(6)). RESULTS WERE REPORTED OUT OF THE LAB. REPEAT TESTING GENERATED LOWER RESULTS AND PATIENT REPORTS WERE AMENDED. THE CUSTOMER NOTIFIED THE PHYSICIANS OF THE FALSE RESULTS AND CONFIRMED THAT NO PATIENT TREATMENT WAS INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO® SYNCHRON CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |