FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO® SYNCHRON CLINICAL SYSTEM

MDR report key: 2111202 · Received June 2, 2011

Report

Report Number
2050012-2011-02071
Event Type
Malfunction
Date Received
June 2, 2011
Date of Event
May 4, 2011
Report Date
May 4, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SAMPLE COLLECTION AND CENTRIFUGATION INFORMATION WERE NOT PROVIDED. CONTROLS RUN BEFORE AND AFTER INCIDENT WERE WITHIN ESTABLISHED RANGES. ON (B)(4) 2011, BCI CUSTOMER TECHNICAL SUPPORT (CTS) DISCUSSED PROPER SAMPLE HANDLING WITH CUSTOMER. CTS ADVISED CUSTOMER TO PERFORM CUVETTE WASH PROCEDURE, HAD CUSTOMER PERFORMED QC AND RAN PRECISION TEST. ON (B)(4) 2011, CUSTOMER REPORTED THAT INSTRUMENT WAS GENERATING MULTIPLE REAGENT SYRINGE MOTION ERROR. ON (B)(4) 2011, FIELD SERVICE ENGINEER (FSE) EVALUATED INSTRUMENT AND FOUND LEAK FROM REAGENT PROBE A, AND PUDDLE OF WATER ON THE COVER BENEATH THE PROBE. FSE REPLACED COMPONENTS TO ADDRESS THIS ISSUE, PERFORMED TOP END ALIGNMENT, AND VERIFIED WASH STATION PERFORMANCE, RAN CARRYOVER PROCEDURE TEST AND RESULTS MET SPECIFICATIONS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT THAT THE UNICEL DXC 600 PRO INSTRUMENT GENERATED TWO (2) FALSE HIGH % HEMOGLOBIN ((B)(6)). RESULTS WERE REPORTED OUT OF THE LAB. REPEAT TESTING GENERATED LOWER RESULTS AND PATIENT REPORTS WERE AMENDED. THE CUSTOMER NOTIFIED THE PHYSICIANS OF THE FALSE RESULTS AND CONFIRMED THAT NO PATIENT TREATMENT WAS INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO® SYNCHRON CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600 PRO NA

Patients

Seq Age Sex Outcome Treatment
1