IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM
Report
- Report Number
- 0002023141-2025-00082
- Event Type
- Injury
- Date Received
- January 9, 2025
- Date of Event
- October 10, 2024
- Report Date
- May 21, 2025
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019508
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). A2: AGE AT TIME OF THE EVENT: UNKNOWN / NOT PROVIDED. G4: PREMARKET IDENTIFICATION: K011028/K013227.
ZIMVIE COMPLAINT NUMBER CMP-(B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT B5: DESCRIBE EVENT OR PROBLEM G3: DATE RECEIVED BY MANUFACTURER G6: TYPE OF REPORT H1: TYPE OF REPORTABLE EVENT H2: FOLLOW UP TYPE H3: DEVICE EVALUATED BY MANUFACTURER H6: ADVERSE EVENT PROBLEM H10: ADDITIONAL NARRATIVE ZIMVIE RECEIVED ONE (1) TSVB10, (IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM), FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED, A FRACTURE HAS BEEN IDENTIFIED ON THE IMPLANTS COLLAR. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1223073. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1223073, FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : FUNCTIONAL : FRACTURE : IMPLANT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS CLINICIAN SELECTS IMPLANT THAT IS UNABLE TO RESIST LONG-TERM OCCLUSAL LOADING. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE FRACTURE HAS BEEN IDENTIFIED ON THE COLLAR. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.
DOCTOR REPORTED IMPLANT COLLAR FRACTURE AT TOOTH #36 AND WAS REMOVED. PAIN WAS REPORTED AS A RESULT OF THE EVENT.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1169302 | IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1223073 | 00889024019508 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male |