FDA Adverse Event Malfunction Summary report: N

METRX® SYSTEM

MDR report key: 21111802 · Received January 9, 2025

Report

Report Number
1030489-2025-00174
Event Type
Malfunction
Date Received
January 9, 2025
Date of Event
November 27, 2024
Report Date
July 28, 2025
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
HRX
UDI-DI
00885074210353
PMA / PMN Number
K002931
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS # (B)(4) - PRODUCT:9560100, LOTNO:781655, THE REQUESTED PHENOMENON WAS OBSERVED AND THE OUTER TUBE WERE DAMAGED DURING THE INCOMING INSPECTION. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT:9560100,LOTNO:781655- DURING THE INCOMING INSPECTION, THE EVENT WAS OBSERVED, AND THAT THE OUTER TUBE HAS SCRATCHES. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, USER FACILITY) REGARDING AN ENDOSCOPE. IT WAS REPORTED THAT SCOPE WAS CLOUDY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1324229 METRX® SYSTEM ARTHROSCOPE HRX MEDTRONIC SOFAMOR DANEK USA, INC 9560100 781655 00885074210353

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown