FDA Adverse Event Injury Summary report: N

GEMINI STONE RETRVL PAIRED WIRE HELICAL BSKT

MDR report key: 2111091 · Received June 2, 2011

Report

Report Number
3005099803-2011-01828
Event Type
Injury
Date Received
June 2, 2011
Date of Event
May 4, 2011
Report Date
May 12, 2011
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
FFL
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED GEMINI STONE RETRIEVAL PAIRED WIRE HELICAL BASKET REVEALED THE DEVICE BASKET WAS DETACHED. THE WIRE SUBASSEMBLY (S/A) DETACHED IN THE MIDDLE OF THE SPLICE CANNULA. THE BASKET WAS BENT. THE WIRE S/A WAS BENT ON THE DISTAL END OF THE BROKEN SPLICE CANNULA. THE OUTER SHEATH WAS KINKED AND BENT. OTHER DEFECTS ON THE DEVICE (BENT BASKET, BENT WIRE S/A, AND KINKED/BENT OUTER SHEATH) INDICATE SIGNIFICANT FORCES WERE APPLIED. IT WAS NOT POSSIBLE TO DETERMINE THE AMOUNT OF FORCE THAT WAS EXERTED ON THE WIRE S/A TO CAUSE IT TO BREAK. THE MOST PROBABLE ROOT CAUSE FOR THIS EVENT IS DETERMINED TO BE UNDETERMINABLE. THE DEVICE HISTORY RECORD REVIEW CONFIRMS THAT THE ACCEPTED DEVICES MET ALL MANUFACTURING SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A GEMINI STONE RETRIEVAL PAIRED WIRE HELICAL BASKET WAS USED IN AN URETEROSCOPY STONE RETRIEVAL PROCEDURE ON A PATIENT (WEIGHT UNKNOWN). ACCORDING TO THE COMPLAINANT, THE BASKET SNAPPED OFF THE DISTAL MANDREL INSIDE THE PATIENT AFTER DURING IT WAS PLACED IN THE PATIENT'S LEFT URETER. THE BASKET WAS RETRIEVED USING GRASPERS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SECOND AGEMINI STONE RETRIEVAL PAIRED WIRE HELICAL BASKET WITHOUT COMPLICATIONS. THE PATIENT IS REPORTED TO BE FINE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A GEMINI STONE RETRIEVAL PAIRED WIRE HELICAL BASKET WAS USED IN AN URETEROSCOPY STONE RETRIEVAL PROCEDURE ON A PATIENT (WEIGHT UNKNOWN). ACCORDING TO THE COMPLAINANT, THE BASKET SNAPPED OFF THE DISTAL MANDREL INSIDE THE PATIENT AFTER DURING IT WAS PLACED IN THE PATIENT'S LEFT URETER. THE BASKET WAS RETRIEVED USING GRASPERS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SECOND AGEMINI STONE RETRIEVAL PAIRED WIRE HELICAL BASKET WITHOUT COMPLICATIONS. THE PATIENT IS REPORTED TO BE FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GEMINI STONE RETRVL PAIRED WIRE HELICAL BSKT DISLODGER, STONE, BASKET, URETERAL, METAL FFL BOSTON SCIENTIFIC - SPENCER M0063301060 14009473

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention