FDA Adverse Event
Malfunction
Summary report: N
AMIS BROACH HANDLE
MDR report key: 2111021
·
Received May 5, 2011
Report
- Report Number
- 3005180920-2011-00024
- Event Type
- Malfunction
- Date Received
- May 5, 2011
- Date of Event
- April 19, 2011
- Report Date
- May 5, 2011
- Manufacturer
- MEDACTA INTERNATIONAL, SA
- Product Code
- JDI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ON THE BASIS OF MEDACTA'S RISK ANALYSIS, THE FAILURE MODE IS UNLIKELY TO CAUSE PATIENT HARM SINCE, IN CASE OF MALFUNCTIONING OF THE BROACH HANDLE, A SECOND BROACH HANDLE IS ALWAYS AVAILABLE IN THE INSTRUMENTATION KIT, PERMITTING THE SURGEON TO COMPLETE THE SURGERY SUCCESSFULLY, AS HAPPENED IN THIS CASE. FROM THE DOCUMENT REVIEW OF THE LOT INVOLVED (084700) NO PARTICULAR ANOMALIES WERE FOUND RELATED TO THE PROBLEM OCCURRED.
Description of Event or Problem · 1
THE MOBILE PEG OF THE HANDLE 01.15.10.0131 WAS BROKEN. NO PATIENT OR USER INVOLVED. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMIS BROACH HANDLE | REUSABLE SURGICAL INSTRUMENT FOR HIP | JDI | MEDACTA INTERNATIONAL, SA | 084700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |