FDA Adverse Event Malfunction Summary report: N

MK1 POOL LIFT

MDR report key: 2110999 · Received April 7, 2011

Report

Report Number
3004468271-2011-00017
Event Type
Malfunction
Date Received
April 7, 2011
Date of Event
March 7, 2011
Report Date
March 10, 2011
Manufacturer
MEDIBO MEDICAL PRODUCTS NV
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE POOL HOIST BOLTS, MAST INC SAFETY CATCH AND BOTTOM OF BOOM AROUND THE FLOOR SOCKET ARE HEAVILY RUSTED. POOL HOIST SUB CHASSIS AND SEAT HAS NORMAL WEAR AND TEAR WITH ONLY SURFACE RUST INSIDE THE SOCKETS OF THE ARM RESTS. PRODUCT NOT UNDER SERVICE CONTRACT.

Description of Event or Problem · 1

CARE GIVER WAS HOISTING RESIDENT OUT OF THE POOL. ONCE OUT OF THE POOL, THE BOOM ARM WAS TURNED AROUND PARALLEL TO THE EDGE OF THE POOL. CARE GIVER THEN DIS-ENGAGED THE CHAIR FROM THE POOL HOIST AND STARTED TO PUSH IT FORWARD, AT THIS POINT THE FRONT LEFT CASTOR (LOOKING FROM THE BACK OF THE SUB CHASSIS CHAIR) FELL OFF CAUSING THE CHAIR TO FALL SIDE WAYS TIPPING RESIDENT OUT. RESIDENT HIT THE SIDE OF THE POOL AS HE FEEL OUT AND ENDED UP BACK IN THE WATER. SWIMMING TEACHER, WHO WAS WALKING AWAY FROM THE HOIST AT THE TIME TURNED AROUND AND JUMPED INTO THE WATER TO RESCUE RESIDENT, WHO WAS THEN TAKEN TO A+E FOR ATTENTION. RESIDENT SUFFERED BRUISING TO HIS CHEST AREA AND RIGHT KNEE, AND ALSO HAD CUTS ON THE MIDDLE OF HIS CHEST, RIGHT ARM AND RIGHT KNUCKLE. DISCHARGED AFTER TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MK1 POOL LIFT HYGIENE AND POOL LIFTERS FSA MEDIBO MEDICAL PRODUCTS NV AX06100.0

Patients

Seq Age Sex Outcome Treatment
1 Other