FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2110904 · Received June 2, 2011

Report

Report Number
6000001-2011-06968
Event Type
Malfunction
Date Received
June 2, 2011
Date of Event
May 9, 2011
Report Date
May 9, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE A FAULTY AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) ON CHANNEL C. TO CORRECT THE CONDITION, THE AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) ON CHANNEL C WAS REPLACED AND CALIBRATED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A COLLEAGUE INFUSION PUMP EXPERIENCED AN AIR-IN-LINE ALARM, AND THIS WAS FOUND TO BE A FALSE ALARM. THIS OCCURRED DURING SERVICE EVALUATION BY BAXTER PERSONNEL. THERE WAS NO KNOWN PATIENT INVOLVED; THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SOFTWARE VERSION IS NOT AVAILABLE AT THIS TIME.