INFUSOR
Report
- Report Number
- 6000001-2011-06963
- Event Type
- Malfunction
- Date Received
- June 2, 2011
- Date of Event
- May 17, 2011
- Report Date
- May 17, 2011
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A LEAK WAS CONFIRMED. UPON SAMPLE RECEIPT, THE DEVICE CONTAINED FLUID IN THE BLADDER. TO VERIFY THE REPORTED CONDITION, THE FILL-PORT CAP WAS REMOVED. UPON REMOVAL, A BACK-FLOW (LEAK) WAS OBSERVED AT THE FILL-PORT. VISUAL EXAMINATION OF THE DISASSEMBLED UNIT REVEALED THE ROOT CAUSE OF THE LEAK WAS A TINY (APPROXIMATELY 0.2 MM IN SIZE) GLASS FRAGMENT UNDER THE CHECK BAND. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT.
(B)(4). THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.
BAXTER (B)(4) RECEIVED A REPORT THAT ONE (1) INFUSOR SV 2.5 DEVICE BACKFLOWED DURING FILLING FROM THE FILLING PORT. THE DEVICE WAS BEING FILLED WITH 5-FU AND SALINE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. THERE WAS NO PATIENT INVOLVEMENT. THE ACTUAL SAMPLE IS AVAILABLE. NO ADDITIONAL INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 10H002 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SALINE| 5-FLUOROURACIL |