FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 2110837
·
Received June 2, 2011
Report
- Report Number
- 2050012-2011-02096
- Event Type
- Malfunction
- Date Received
- June 2, 2011
- Date of Event
- May 2, 2011
- Report Date
- May 2, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER TECHNICAL SUPPORT (CTS) INSTRUCTED THE CUSTOMER TO TAKE LINE 36 OFF AND OBSERVED FLUID COMING OUT OF THE DRAIN PORT. NO OBSTRUCTION WAS NOTED. CTS HAD CUSTOMER PRIME AND IT'S STILL FILLING UP DRAIN.(B)(4) A BCI FIELD SERVICE ENGINEER (BCI) FOUND J2 VALVE ON ISE DRAIN ASSEMBLY NOT WORKING. FSE SHUT DOWN AND RESTARTED THE INSTRUMENT; HOWEVER, THE VALVE WAS STILL NOT OPERATIONAL. FSE REPLACED J2 ON VALVE ASSEMBLY, PRIMED ISE, CALIBRATED ISE AND RAN QC. THIS ISSUE HAS BEEN RESOLVED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT GETTING A CAL REFERENCE DRIFTS AND THE FLOW CELL DRAIN OVERFLOWED. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |