FDA Adverse Event Malfunction Summary report: N

FOXCROSS PTA CATHETER

MDR report key: 2110750 · Received June 1, 2011

Report

Report Number
2024168-2011-03826
Event Type
Malfunction
Date Received
June 1, 2011
Date of Event
May 7, 2011
Report Date
May 10, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
LIT
PMA / PMN Number
K102705
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. BALLOON MATERIAL RUPTURES CAN BE AFFECTED BY NUMEROUS FACTORS INCLUDING, BUT IS NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, MATERIALS, INTERACTIONS WITH OTHER DEVICES, A PREVIOUSLY IMPLANTED STENT, LESION CALCIFICATION AND TORTUOSITY, OR INSUFFICIENT PREPARATION PRIOR TO USE. TO ENSURE THIS TYPE OF DAMAGE IS NOT A RESULT OF A POTENTIAL MANUFACTURING RELATED DEFICIENCY, ALL DILATATION CATHETERS ARE VISUALLY INSPECTED AND LEAK TESTED DURING THE MANUFACTURING PROCESS. A SAMPLING OF UNITS IS ALSO DESTRUCTIVELY TESTED TO VERIFY RATED BURST PRESSURE AND BALLOON INTEGRITY. A CONCLUSIVE CAUSE FOR THE BALLOON RUPTURE COULD NOT BE DETERMINED; HOWEVER, THERE IS NO INDICATION TO SUGGEST A PRODUCT QUALITY DEFICIENCY. A REVIEW OF THE FINISHED DEVICE LOT HISTORY RECORDS DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT AND THERE HAVE BEEN NO OTHER INCIDENTS REPORTED FOR THIS LOT. ADDITIONALLY, A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS PERFORMED AND THERE HAVE BEEN NO OTHER INCIDENTS REPORTED FOR THIS LOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION OF THE FOXCROSS DILATATION CATHETER, THE BALLOON WOULD NOT INFLATE. A PINHOLE IN THE BALLOON WAS SUSPECTED. THE DEVICE WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. ANOTHER SAME DEVICE WAS USED TO PERFORM DILATATION SUCCESSFULLY. THERE WAS NO SIGNIFICANT CLINICAL DELAY IN THE PROCEDURE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FOXCROSS PTA CATHETER PERIPHERAL DILATATION CATHETER LIT AV-TEMECULA-CT 692730

Patients

Seq Age Sex Outcome Treatment
1