XPERT SELF-EXPANDING STENT SYSTEM
Report
- Report Number
- 2024168-2011-03823
- Event Type
- Malfunction
- Date Received
- June 1, 2011
- Date of Event
- May 11, 2011
- Report Date
- May 11, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- FGE
- PMA / PMN Number
- K050534
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. POTENTIAL FACTORS WHICH COULD CONTRIBUTE PREMATURE DEPLOYMENT PRIOR TO USE INCLUDE, BUT ARE NOT LIMITED TO, MANUFACTURING, HANDLING, INSUFFICIENT SUPPORT DURING PREP, KINKS IN THE SHAFT, LOOSENING OF THE TUOHY BORST VALVE, INTERACTION DURING LOADING ONTO A GUIDE WIRE, OR INTERACTIONS WITH THE INTRODUCER SHEATH AND/OR ASSOCIATED DEVICES DURING INSERTION. TO ENSURE THIS IS NOT A RESULT OF A MANUFACTURING DEFICIENCY, ALL SELF EXPANDING STENT DELIVERY SYSTEMS ARE INSPECTED FOR DAMAGE DURING THE MANUFACTURING PROCESS. A REVIEW OF THE FINISHED DEVICE LOT HISTORY RECORDS DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS FOR THIS LOT AND THERE HAVE BEEN NO OTHER INCIDENTS REPORTED FOR THIS LOT. WITHOUT HAVING THE DEVICE TO EXAMINE, A CONCLUSIVE CAUSE COULD NOT BE DETERMINED; HOWEVER, THERE IS NO INDICATION TO SUGGEST A PRODUCT QUALITY DEFICIENCY.
IT WAS REPORTED THAT DURING PREPARATION OF THE XPERT STENT WITH SALINE WATER, THE PART OF THE STENT NEAR THE TIP OF THE DELIVERY DEVICE OPENED 5MM. THE DEVICE WAS NOT USED IN THE PATIENT AND THERE WAS NO PATIENT INVOLVEMENT. ANOTHER UNKNOWN DEVICE WAS USED TO COMPLETE THE PROCEDURE. THOUGH REQUESTED, THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | XPERT SELF-EXPANDING STENT SYSTEM | SELF-EXPANDING STENT AND DELIVERY SYSTEM | FGE | AV-TEMECULA-CT | 700704 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |