FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000 SR ANALYZER

MDR report key: 2110706 · Received June 1, 2011

Report

Report Number
1628664-2011-00249
Event Type
Malfunction
Date Received
June 1, 2011
Report Date
May 10, 2011
Manufacturer
ABBOTT MANUFACTURING INC
Product Code
JJE
PMA / PMN Number
K983212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REFER TO RESULTS OF INVESTIGATION BELOW. AN EVALUATION WAS PERFORMED BY REVIEWING THE COMPLAINT TEXT, INSTRUMENT SERVICE HISTORY, AND OTHER CUSTOMER COMPLAINTS FOR SIMILAR ISSUES. THE CUSTOMER'S ISSUE WAS RESOLVED BY REPLACING THE STAT PIPETTOR PROBE AND PERFORMING THE SAMPLE PIPETTOR CALIBRATION. TROUBLESHOOTING AND DIAGNOSTICS, OBSERVED PROBLEMS, ERRATIC ASSAY RESULTS IN THE ARCHITECT SYSTEM OPERATIONS MANUAL LISTS MULTIPLE PROBABLE CAUSES FOR ERRATIC RESULTS, INCLUDING PROBE TIP BENT OR DAMAGED AND THE PROBE NOT POSITIONED CORRECTLY. CORRECTIVE ACTIONS INCLUDE REPLACING THE PROBE AND PERFORMING PIPETTOR CALIBRATION. A REVIEW OF THE SERVICE HISTORY FOR ARCHITECT SERIAL # (B)(4) DID NOT IDENTIFY ISSUES RELATED TO THIS COMPLAINT. A REVIEW OF CUSTOMER COMPLAINTS AND QUALITY METRICS DID NOT IDENTIFY AN ADVERSE TREND RELATED TO THE CUSTOMER'S ISSUE. BASED ON THE AVAILABLE INFORMATION, THE LIKELY CAUSE OF THE ISSUE WAS A DAMAGED STAT PROBE OR POOR PROBE ALIGNMENT AND WAS RESOLVED BY REPLACING THE PROBE AND CALIBRATING THE PIPETTOR. NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE ACCOUNT STATED THAT THE ARCHITECT ANALYZER GENERATED AN INITIAL TROPONIN RESULT OF 0.033 NG/ML (B)(6). THE SAMPLE WAS REPEATED WITH A RESULT OF 0.010 NG/ML (NEGATIVE). THE ACCOUNT WAS USING A CUTOFF OF 0.03 NG/ML AS (B)(6). THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000 SR ANALYZER AUTOMATED IMMUNOASSAY ANALYZER JJE ABBOTT MANUFACTURING INC

Patients

Seq Age Sex Outcome Treatment
1 LOT # 60180UN10| ARCHITECT TROPONIN LN 2K41-20 LOT 60180UN10| ARCHITECT STAT TROPONIN-I LIST # 2K41-20