FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2110556 · Received June 1, 2011

Report

Report Number
1423500-2011-06913
Event Type
Malfunction
Date Received
June 1, 2011
Date of Event
May 17, 2011
Report Date
May 17, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING INVESTIGATION BY BAXTER, THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE/USER ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR (SE) 2240 (AIR IN SET) OCCURRED DURING INITIAL DRAIN WAS NOT CONFIRMED; HOWEVER PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE CAUSED BY THE PATIENT NOT CONNECTED WHEN THERAPY INITIATED. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A LABELING REVIEW FOUND THE HOME CHOICE USER MANUAL TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).

Description of Event or Problem · 1

A HOME PATIENT (HP) CONTACTED (B)(4) REGARDING A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOME CHOICE (HC) UNIT DURING INITIAL DRAIN. THE HP STATED THEY PRESSED GO BEFORE CONNECTING THEN CONNECTED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP WITH CLEARING THE ALARM THEN EXPLAINED A SE 2240 INDICATES A LARGE AMOUNT OF AIR HAS ENTERED SETUP AND THAT THE HP NEEDED TO END THERAPY AND START OVER WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. A FOLLOW UP WAS DONE VIA PHONE CALL. THE HP CONFIRMED THEY STARTED OVER WITH NEW SUPPLIES. THE HP HAS CONTINUED THERAPY. NO INJURY OR MEDICAL INTERVENTION REPORTED AS A RESULT OF THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 67 YR HOME CHOICE