INRATIO
Report
- Report Number
- 2027969-2011-00916
- Event Type
- Malfunction
- Date Received
- April 25, 2011
- Date of Event
- March 24, 2011
- Report Date
- April 25, 2011
- Manufacturer
- ALERE SAN DIEGO, INC
- Product Code
- GJS
- PMA / PMN Number
- 021923
- Removal / Correction Number
- NI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DATA ANALYSIS WAS NOT PERFORMED ON CUSTOMER'S RESULTS SINCE INRATIO AND REFERENCE TESTS WERE DONE FOUR HOURS APART. SINCE TIME BETWEEN TESTS EXCEEDED THREE HOURS, THERE IS A HIGH POSSIBILITY THAT THE DISCREPANCIES ARE DUE TO CHANGES IN THE STATUS OF THE PT. THREE HOURS IS CONSIDERED A REASONABLE LENGTH OF TIME BETWEEN TWO READINGS IN ORDER FOR A COMPARISON TO BE VALID. NO PRODUCT IS EXPECTED TO BE RETURNED. RETAINED STRIP TESTING FROM A PREVIOUS CASE REVEALED THAT TEST RESULT COMPARISONS MET ACCURACY CRITERIA. AS OF 4/25/2011, (B)(4) COMPLAINTS HAVE BEEN REPORTED FOR LOT #243934. THIS REACHES THE ACTION THRESHOLD. BRICK LOT NUMBER 233254 HAS STRIP CODE (B)(4), MATERIAL WAS SPLIT INTO TWO DIFFERENT LOTS: LOT #243934 INTO 12 PACKS (SMALLER LOT) AND LOT #241836 INTO 12 PACKS (LARGER LOT). THEREFORE, (B)(4) COMPLAINTS HAVE BEEN REPORTED FOR LOT NUMBERS 243934 AND 241836. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLD ((B)(4)), CORRECTIVE ACTION IS NOT REQUIRED AT THIS TIME. (B)(4) 2010 TO (B)(4) 2011, 9 THERAPEUTIC AND 3 NORMAL DONOR SAMPLE TESTS HAVE BEEN PERFORMED. TEST RECORDS INDICATED THAT ALL STRIP TEST RESULTS MET PRODUCT PERFORMANCE WHEN COMPARED TO THE RESULTS FROM IN VIVO TESTS. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED. INVESTIGATION RESULTS FROM A PREVIOUS CASE: THE ALLOWABLE DIFFERENCE BETWEEN THE INRATIO INR'S AND SYSMEX INR'S WAS CALCULATED. AT LEAST 2 OUT OF 3 REPLICATES FOR BOTH SAMPLES TESTED ON STRIP LOT #243934 ARE WITHIN THE ALLOWABLE BIAS (+ OR -1.0). NO DISCREPANT RESULTS WERE PRODUCED ON RETURNED AND IN-HOUSE METERS. THESE MEET THE ABOVE CRITERIA. NO FURTHER ACTION IS REQUIRED.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B)(6) 2011, INRATIO: 2.4, LAB: 1.8. PT'S THERAPEUTIC RANGE: 2.0 - 3.0 INR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC | 100071 | 243934 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |