FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2110463
·
Received April 25, 2011
Report
- Report Number
- 2531779-2011-02801
- Event Type
- Malfunction
- Date Received
- April 25, 2011
- Report Date
- March 28, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. DURING EVAL, THE BOLUS BUTTON WAS FOUND TO BE MISSING.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD IS RESPONDING INTERMITTENTLY OVER THE PAST 2 MONTHS. IT WAS REPORTED THAT THE PUMP DOES NOT GET WET AND THERE IS NO DAMAGE TO THE KEYPAD. THE PT CONTINUES TO USE THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |