FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2110463 · Received April 25, 2011

Report

Report Number
2531779-2011-02801
Event Type
Malfunction
Date Received
April 25, 2011
Report Date
March 28, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. DURING EVAL, THE BOLUS BUTTON WAS FOUND TO BE MISSING.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD IS RESPONDING INTERMITTENTLY OVER THE PAST 2 MONTHS. IT WAS REPORTED THAT THE PUMP DOES NOT GET WET AND THERE IS NO DAMAGE TO THE KEYPAD. THE PT CONTINUES TO USE THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1