ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2110461
·
Received April 25, 2011
Report
- Report Number
- 2531779-2011-02796
- Event Type
- Malfunction
- Date Received
- April 25, 2011
- Report Date
- March 25, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. ALL KEYPAD BUTTONS DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. IT WAS ALSO OBSERVED THAT MULTIPLE BUTTON PRESSES WERE REQUIRED TO ENGAGE A RESPONSE. ALL BUTTONS DID NOT HAVE NORMAL SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD REQUIRED MULTIPLE PRESSES TO RESPOND. THE PT REPORTED HE DOES NOT GET THE PUMP WET.