MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL
Report
- Report Number
- 2032227-2025-105142
- Event Type
- Malfunction
- Date Received
- January 8, 2025
- Date of Event
- December 28, 2024
- Report Date
- February 27, 2025
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000545604
- PMA / PMN Number
- P150001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- 003
Narratives
PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. PUMP SUCCESSFULLY DOWNLOADED TRACES AND HISTORY FILE USING THUMP. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM TRIGGERED BY THREE CONSECUTIVE PUMP ERROR 63 ALARMS ON (B)(6) 2024 23:29:45.000, (B)(6) 2024 23:29:53.000, (B)(6) 2024 23:30:06.000, (B)(6) 2024 23:30:17.000 HARDWAREERRORALARM (63) FILE NUMBER: 2017, LINE NUMBER: 147 SW/HW: HW (POSSIBLE HW) GROUPING: TWI INTERFACE ON MAIN/MOTOR PROCESSOR. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND CORROSION ON THE ELECTRONIC ASSEMBLY. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE ALSO NOTED DURING VISUAL INSPECTION: PILLOWING KEYPAD OVERLAY, CRACKED KEYPAD OVERLAY, SERIAL NUMBER LABEL MISSING AND CRACKED CASE (BATTERY TUBE). CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO THREE CONSECUTIVE PUMP ERROR 63 ALARMS. PUMP ERROR 63 ALARM, SUSPECTED ON HW. PUMP EXPOSED TO MOISTURE CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED CRITICAL PUMP ERROR (OPEN BOOK IMAGE), PUMP ERROR 63(HARDWARE LOW LEVEL FAILURES). THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1880. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1880 WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1105906 | MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1880 | NG3115820H | 000000763000545604 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | Male |