FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2110355
·
Received April 21, 2011
Report
- Report Number
- 3002158293-2011-00465
- Event Type
- Malfunction
- Date Received
- April 21, 2011
- Date of Event
- February 26, 2011
- Report Date
- April 21, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204/COMMUNICATION ERRORS) HAS BEEN CONFIRMED. AS RECEIVED, THE TRUNK CABLE CONNECTOR WAS BROKEN. THE ROOT CAUSE OF THE BROKEN CONNECTOR CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. ONCE THE TRUNK WAS REPLACED, THE BELT WAS FULLY FUNCTIONAL. NO ADVERSE EVENT RESULTED FROM THE BROKEN CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
ZOLL CUSTOMER SUPPORT REVIEWED THE DOWNLOAD OF A (B)(6) FEMALE PATIENT WHICH REVEALED SEVERAL SERVICE CODE 204S AND BELT COMMUNICATION ERRORS. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |