FDA Adverse Event Malfunction Summary report: N

SECURE 3 MED/SURG BED

MDR report key: 2110333 · Received May 20, 2011

Report

Report Number
1831750-2011-04921
Event Type
Malfunction
Date Received
May 20, 2011
Date of Event
May 9, 2011
Report Date
May 9, 2011
Manufacturer
STRYKER MEDICAL
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

RESULT: LOAD CELL.

Description of Event or Problem · 1

IT WAS REPORTED BY SERVICE REPORT THAT THE BED SCALE, I-BED, AND BED EXIT ARE NOT WORKING. THE FOOT RIGHT LOAD CELL IS READING ZERO. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES ARE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SECURE 3 MED/SURG BED A/C POWERED HOSPITAL BED FNL STRYKER MEDICAL 3002S3EX NA

Patients

Seq Age Sex Outcome Treatment
1