PROXIMATE ILS CURVED CIR STAPL
Report
- Report Number
- 3005075853-2011-02223
- Event Type
- Malfunction
- Date Received
- June 1, 2011
- Date of Event
- May 12, 2011
- Report Date
- May 13, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). UNCUT WASHER - BLEMISHED. ADDITIONAL INFORMATION REQUESTED. THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION WITHOUT STAPLES PRESENT AND WITH THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT BEFORE FIRING THE DEVICE, THE ORANGE INDICATOR SHOULD BE FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE. IN ADDITION, IT SHOULD BE NOTED THAT IF THE FIRING SEQUENCE IS NOT COMPLETE (PLASTIC TO PLASTIC), STAPLES COULD BE PARTIALLY DEPLOYED WITHOUT CUTTING THE WASHER. FOR MORE INFORMATION, PLEASE REFER TO THE INSTRUCTIONS FOR USE. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AFTER THE SURGEON FIRED THE DEVICE, THE STAPLES WERE MALFORMED AND FELL OFF. THE SURGEON USED HAND SEWING TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE ILS CURVED CIR STAPL | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4T51H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |