FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM

MDR report key: 21102874 · Received January 8, 2025

Report

Report Number
0002023141-2025-00067
Event Type
Injury
Date Received
January 8, 2025
Date of Event
December 2, 2024
Report Date
May 13, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019515
PMA / PMN Number
K061410
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). G4: PREMARKET IDENTIFICATION: K013227.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). ZIMVIE RECEIVED ONE (1) TSVB11, (IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM), FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED, A FRACTURE HAS BEEN IDENTIFIED ON THE IMPLANTS COLLAR. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1230026. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1230026, FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : FUNCTIONAL : FRACTURE : IMPLANT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS CLINICIAN SELECTS IMPLANT THAT IS UNABLE TO RESIST LONG-TERM OCCLUSAL LOADING. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE FRACTURE HAS BEEN IDENTIFIED ON THE COLLAR. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED A FRACTURE OF IMPLANT COLLAR AT TOOTH SITE #46 DISCOVERED AFTER SEVERAL UNSCREWINGS AND WAS REMOVED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
883113 IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL 1230026 00889024019515

Patients

Seq Age Sex Outcome Treatment
1 53 YR Male