FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2110146 · Received May 19, 2011

Report

Report Number
1824206-2011-02780
Event Type
Malfunction
Date Received
May 19, 2011
Date of Event
April 27, 2011
Report Date
April 27, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH ISOLATED THE ISSUE TO A LEAKING HEAD HI/LOW UP VALVE. HE TIGHTENED THE VALVE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE BED IS DRIFTING DOWN INTO THE TREND POSITION IN ABOUT 20 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1