FDA Adverse Event Malfunction Summary report: N

S3 SHOULDER PLATE 11 HOLE RIGHT

MDR report key: 2110030 · Received May 17, 2011

Report

Report Number
1818910-2011-08800
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 18, 2011
Report Date
April 18, 2011
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXT
PMA / PMN Number
K060290
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE LOT CODE REQUIRED FOR THE PLATE AND THE PART/LOT CODES REQUIRED FOR THE SCREWS WERE NOT PROVIDED. A PREVIOUS INVESTIGATION FOUND ALTHOUGH THE COMPLAINT IS NON-VERIFIABLE, THERE ARE PREVIOUS COMPLAINTS FOR THIS FAILURE MODE. PT ANATOMY AND SURGICAL TECHNIQUE CAN IMPACT THE PEG ENGAGEMENT EVEN IF THE PRODUCT IS WITHIN SPEC. THIS IS DISCUSSED IN DVA-104443-DFMEA AND DVA-10443-RMR AND IS APPROVED WITH AN IFR (INVESTIGATE FURTHER REDUCTION) DESIGNATION. THE ENGAGEMENT ISSUES ARE IMPACTED BY PT ANATOMY AND SURGICAL TECHNIQUE. ALTHOUGH THE RISK IS CONSIDERED ACCEPTABLE PER THE RMR (RISK MANAGEMENT REPORT), PRODUCT DEVELOPMENT, QUALITY AND MARKETING ARE INVESTIGATION WAYS TO FURTHER REDUCE THE OCCURRENCE OF PEG ENGAGEMENT ISSUES. (B)(4) WAS OPENED ON (B)(4) 2011 TO DETERMINE IF ANY ADD'L ACTIONS MAY DECREASE THE LIKELIHOOD OF PEG ENGAGEMENT ISSUES. ANY ROOT CAUSE AND/OR CORRECTIVE ACTIONS WILL BE DOCUMENTED IN (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ANY ADD'L INFO BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

SCREWS WOULD NOT LOCK INTO THE PLATE. THIS RESULTED IN A SURGICAL DELAY OF 30-45 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 S3 SHOULDER PLATE 11 HOLE RIGHT 87 LXT LXT DEPUY ORTHOPAEDICS, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 79 YR