FDA Adverse Event Malfunction Summary report: N

GSOURCE, LLC

MDR report key: 21097821 · Received January 7, 2025

Report

Report Number
2249529-2025-00001
Event Type
Malfunction
Date Received
January 7, 2025
Date of Event
December 9, 2024
Report Date
January 7, 2025
Manufacturer
GSOURCE, LLC
Product Code
HTW
UDI-DI
00840314246752
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AS OF 01/07/2025, NO BROKEN PART HAS BEEN RECEIVED BY GSOURCE FROM HOSPITAL. LOT# OF AFFECTED 02 PIECES WAS 5232442 AS PER EMAIL CHAIN FROM HOSPITAL. NO IMAGE OF THE BROKEN PART WAS SHARED WITH GSOURCE. THE EMAIL BRIEFLY MENTIONED BOTH INCIDENTS OF BREAKING DRILL BITS AND REMAINING IN PATIENT'S BODY, HOWEVER, IT DID NOT PROVIDE ANY SPECIFIC DETAILS ABOUT THE TYPE OF SURGERY PERFORMED. INSPECTION DATA SHEET OF GS 86.8427 FOR LN: 5232442 WAS REVIEWED AND FOUND TO BE CONFORMING. THERE WERE (B)(4) PCS IN BATCH WITH LN: 5232442. SEE DHR TRANSACTIONS SNIP WHICH INDICATES THAT (B)(4) PCS AT MOST WERE SHIPPED TO (B)(6) SURGICAL CENTER WITH LN: 5232445, (B)(4) PCS WERE IN STOCK AND ARE CURRENTLY QUARANTINED, THE REST WERE SOLD TO CUSTOMER(S) WITH NO REGISTERED COMPLAINTS. IN THE LAST 10 YEARS, NO COMPLAINT OF BREAKING GS 86.8427 WAS LOGGED ON COMPLAINT LOG. THERE WAS NO SIGNIFICANT DEVIATION IN SELLING QUANTITY OF DRILLS FROM 2023 TO 2024. MANUFACTURING PAPERWORK REVIEWED (CERTIFICATE OF MATERIAL, HEAT TREAT, ETC.) AND FOUND TO BE CONFORMING TO SPECIFICATIONS. RAW MATERIAL WAS PURCHASED FROM EURE INOX FOR LN: 5232442, HOWEVER ALL OTHER MATERIAL CERTIFICATES WITHIN DHR INDICATE KOLB. CHANGING RAW MATERIAL SOURCE DOES NOT HINT AS A ROOT CAUSE BECAUSE THE PREVIOUS BREAKAGE COMPLAINTS HAD KOLB AS RAW MATERIAL PROVIDER. THE WEIGHT OF DRILL BIT WITH LN: 5232442 WEIGHS 4.5 G WHEREAS INSPECTION SAMPLE WEIGHS 5.0 G HENCE NO SIGNIFICANT DEVIATION. THE HARDNESS IS ALSO IN RANGE HENCE COULD NOT VERIFY CUSTOMER'S CLAIM ABOUT "DR THOUGHT THE DRILL BIT LOOKED A LITTLE DIFFERENT AND FELT DIFFERENT." GSOURCE HAS QUARANTINED (B)(4) PCS IN STOCK OF GS 86.8427 WHICH INCLUDES LN[?] 5232442(08), 5232450(70), 5232437(16), 5232431(5), 5232441(20), 5232440(72). IN-HOUSE TESTING CONDUCTED. THE LITERATURE IDENTIFIES DRILL BITS AS SURGICAL EQUIPMENT THAT COMMONLY BREAKS DURING USE. IF A BROKEN PART IS RECEIVED, THEN MORE DETAILS WILL BE ADDED TO THE COMPLAINT FILES AND INCLUDED TO FOLLOW UP MEDWATCH FORM.

Description of Event or Problem · 0

TIP OF DRILL BIT BROKE DURING SURGERY AND LEFT IN PATIENT BODY SINCE REMOVING WOULD CAUSE MORE HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
870372 GSOURCE, LLC TWIST DRILL 5" STR SHANK DOUBLE FLUTED Ø 2.7MM 30MM ROUND END HTW GSOURCE, LLC 5232442 00840314246752

Patients

Seq Age Sex Outcome Treatment
1 55 YR Male Other