UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2011-02055
- Event Type
- Malfunction
- Date Received
- June 1, 2011
- Date of Event
- May 3, 2011
- Report Date
- May 4, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
NO SPECIFIC SAMPLE INFORMATION WAS PROVIDED. THE CUSTOMER REPORTED THAT CALIBRATION WAS FAILING AND OTHER CHEMISTRIES WERE HAVING PRECISION ISSUES, AND THAT QC WAS OUT OF RANGE. NO OTHER CHEMISTRY RESULTS WERE PROVIDED. SERVICE VISITED ON (B)(4) 2011, AND THE FIELD SERVICE ENGINEER (FSE) OBSERVED IMPRECISION ON TG CONTROLS AND SAMPLES. IT WAS FOUND THAT ON THE (B)(4) THE CUSTOMER HAD RUN THE MEDICA WATER SYSTEM DRY PULLING OFF WATER. THIS ALLOWED BEADS TO BE INTRODUCED INTO THE SYSTEM. BOTH MIXER WASH STATIONS WERE PLUGGED AND CUVETTE WASH STATION PROBE 2 WAS ALSO PLUGGED. THIS CAUSED CARRYOVER. THE FSE CLEANED MIXER WASH STATION INSERTS AND REPLACED CUVETTE WASH PROBE 2. THE FSE FLUSHED MIXER STATIONS AND CLEANED HYDRO DI WATER CANISTERS. SIEMENS WATER CAME IN AND PLACED A 1 MICRON FILTER ON THE OUTPUT OF THE MEDICA TO PREVENT FURTHER ISSUES. THE FSE PERFORMED TRIG/URIC CARRYOVER TEST, WHICH PRODUCED RESULTS WITHIN THE PUBLISHED SPECIFICATIONS. THE FSE ALSO RAN 20 POINT PRECISIONS ON TG, BUN, LDH AND AMM. ALL TESTS PRODUCED RESULTS WITHIN THE PUBLISHED SPECIFICATIONS. SERVICE VISITED AGAIN ON (B)(4) 2011, AND REPLACED CLEANED MIXER WASH STATION INSERTS. THE FSE RAN TRIG/URIC CARRYOVER, WHICH PRODUCED RESULTS WITHIN THE SPECIFICATIONS. THE SYSTEM WAS OPERATIONAL, AND AS OF (B)(4) 2011, THERE HAS BEEN NO FURTHER SERVICE CALLS FOR THIS ISSUE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRATIC TRIGLYCERIDES (TG) RESULTS GENERATED BY UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER FOR MULTIPLE PATIENTS. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. REPEAT TESTING PRODUCED DISCREPANT RESULTS. THE CUSTOMER INDICATED THAT THERE WAS NO EFFECT TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMSTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |