FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK

MDR report key: 21096664 · Received January 7, 2025

Report

Report Number
1213809-2024-01038
Event Type
Malfunction
Date Received
January 7, 2025
Date of Event
November 25, 2024
Report Date
January 16, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

(B)(6). FOLLOW UP MDR FOR DEVICE EVALUATION. THREE PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. TWO PHOTO SHOWS DIFFERENT CLOSE-UP VIEWS OF A FULLY ASSEMBLED, LOOSE SYRINGE WITH A COMPLETELY BROKEN COLLAR, COMPROMISING THE SYRINGE'S INTEGRITY. THE THIRD IMAGE PARTIALLY DEPICTS A DAMAGED BARREL WITH A BROKEN COLLAR AND A SECTION OF A NEEDLE HUB PROTECTED BY A SHIELD. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATION. THE POTENTIAL ROOT CAUSE FOR THE DAMAGED BARREL COULD NOT BE DETERMINED FROM THE PHOTO PROVIDED. A PHYSICAL SAMPLE IS REQUIRED FOR A MORE THOROUGH EVALUATION AND POTENTIAL ROOT CAUSE DETERMINATION. THE LOT NUMBER IS UNKNOWN, THEREFORE DEVICE HISTORY RECORD REVIEW (DHR) OR QUALITY NOTIFICATION REVIEW (QN) COULD NOT BE PERFORMED.

Description of Event or Problem · 0

MATERIAL # 309628. BATCH # 3122722. IT WAS REPORTED THAT THE BD LUER-LOK LUER WAS CRACKED / DAMAGED / DEFORMED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. ¿THE REPORTER STATED THAT THE "THE TIP OF THE SYRINGE BROKE OFF" DURING PREPARATION¿. CATALOG #: 309628. LOT #: 3122722.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1321411 BD LUER-LOK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 3122722 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown