FDA Adverse Event Malfunction Summary report: N

CERTAIN GOLD-TITE TM HEXED SCREW

MDR report key: 21096247 · Received January 7, 2025

Report

Report Number
0001038806-2025-00044
Event Type
Malfunction
Date Received
January 7, 2025
Date of Event
August 11, 2024
Report Date
May 13, 2025
Manufacturer
BIOMET 3I
Product Code
NHA
UDI-DI
00844868008538
PMA / PMN Number
K072642
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4).

Additional Manufacturer Narrative · 0

ZIMVIE DID NOT RECEIVE ONE (1) IUNIHG, (CERTAIN GOLD-TITE TM HEXED SCREW) FOR EVALUATION. VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM NUMBER, DEVICE HISTORY RECORD (DHR) REVIEW, AND RISK MANAGEMENT FILE (RMF). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1251185. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1251185 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿DENTAL : FUNCTIONAL : FRACTURE : SCREW.¿ THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW AS PER (B)(4), THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION WAS NOT ESTABLISHED. WITHOUT DEVICE RECEIPT, THE REPORTED EVENT IS NON-VERIFIABLE. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT, G3: DATE RECEIVED BY MANUFACTURER, G6: CHECKED "FOLLOW-UP," H6: ENTERED EVALUATION CODES, H10: ADDED MANUFACTURER NARRATIVE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE HEX OF THE ABUTMENT AND SCREW FRACTURED, NOT THE IMPLANT. THE IMPLANT WAS NOT FRACTURED AND WAS NOT REMOVED. THE BROKEN SCREW WAS RETRIEVED, AND THE LAB WILL BE REMAKING THE ABUTMENT SYMPTOMS AS A RESULT OF THE EVENT: PAIN & INFLAMMATION.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1012345 CERTAIN GOLD-TITE TM HEXED SCREW DENTAL SCREW NHA BIOMET 3I 1251185 00844868008538

Patients

Seq Age Sex Outcome Treatment
1 56 YR Male