NOA MEDICAL INDUSTRIES
Report
- Report Number
- 1931947-2023-00002
- Event Type
- Malfunction
- Date Received
- January 7, 2025
- Date of Event
- December 5, 2023
- Report Date
- October 28, 2024
- Manufacturer
- NOA MEDICAL INDUSTRIES
- Product Code
- FNL
- UDI-DI
- 00817492012745
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE LATCH IS A TWO STEP PROCESS WHERE THE BUTTON MUST BE PUSHED IN AND THE SIDE RAIL MOVED. THERE IS A POSSIBILITY THAT THEY DIDN'T GET THE SIDE RAIL PUSHED ALL THE WAY UP TO CAUSE THE BUTTON TO POP OUT AND THE RAIL TO LOCK INTO PLACE. A REPLACEMENT WAS SENT TO THE CUSTOMER. WE WERE WAITING FOR THE PART TO RETUN INCLUDING CONTACTING TWICE AND SENDING A RETURN LABEL. DEVICE WAS NOT RETURNED FOR INSPECTION SO A MALFUNCTION COULD NOT BE DETERMINED. IN HOUSE INVESTIGATION FOUND A SLIGHT TOLERANCE ISSUE ON THE FRAMES THAT ALLOWED THE LATCH ATTACHMENT PINS TO MOVE, CATCHING THE BUTTON ON THE COVER AND EITHER BREAKING IT OR HANGING UP ON THE COVER AND GETTING STUCK. BUTTON WAS ATTACHED TO THE LATCH WITH PLASTIC PINS THAT BROKE EASILY. THE RED BUTTON WAS REDESIGNED TO SHORTEN AND BE HELD IN PLACE WITH SCREWS INSTEAD OF PINS. CHANGED LATCH ATTACHMENT PINS TO A THREADED ROD AND PLATES WHICH SECURES THE LATCH TO PREVENT MOVEMENT. PIN LENGTH ON THE BACK OF THE LATCH WAS ALSO FOUND TO AFFECT PERFORMANCE. A CORRECTION IS PLANNED TO REPLACE FRAME ASSEMBLY (FRAME, LATCH, RED BUTTON).
(B)(4), ON TUE, DEC 5, 2023 AT 9:41AM WE RECEIVED AN EMAIL ALLEGING THAT THE RED BUTTON ON THE SIDE RAIL THAT RELEASES THE RAIL TO DROP DOWN WAS STUCK INSIDE AND DOESN'T POP OUT AS IT SHOULD TO LOCK THE RAIL IN PLACE. THE RAIL ISN'T LOCKING INTO PLACE, WHICH ALLOWED HIM TO FALL OUT OF HIS BED DUE TO THE RAIL DROPPING FROM THE UPRIGHT POSITION. FORTUNATELY, HE IS OKAY BUT THE RAIL IS NOT. INCIDENT WAS REPORTED FROM A USER VIA A DISTRIBUTOR, NO OTHER INFORMATION WAS PROVIDED. WE SENT A REPLACEMENT WITH A RETURN LABEL SO THE DEVICE COULD RETURN FOR INVESTIGATION. WE REACHED OUT AGAIN LATER WHEN THE PART HAD NOT RETURNED OFFERING TO SEND ANOTHER RETURN LABEL BUT NEVER HEARD BACK FROM THE CUSTOMER. WE COULD NOT DETERMINE THAT A MALFUNCTION OCCURED. REPORT IS BEING FILED AFTER THE 30 DAY TIMELINE DUE TO CONFUSION OVER THE FILING OF MDRS WITH LITTLE INFORMATION. WE ARE REPORTING WITH THE INFORMATION REASONABLY KNOWN TO US OUT OF AN ABUNDANCE OF CAUTION. SIDE RAIL IS PART OF THE BED PACKAGE. THERE ARE NO PROBLEMS ASSOCIATED WITH THE BED ITSELF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 868974 | NOA MEDICAL INDUSTRIES | SIDE RAIL SET - FOOT BEHAVIORAL HEALTH BED (GRAY) | FNL | NOA MEDICAL INDUSTRIES | 7540228LGRY | 00817492012745 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male | Other |