FDA Adverse Event Malfunction Summary report: N

NOA MEDICAL INDUSTRIES

MDR report key: 21091763 · Received January 7, 2025

Report

Report Number
1931947-2023-00002
Event Type
Malfunction
Date Received
January 7, 2025
Date of Event
December 5, 2023
Report Date
October 28, 2024
Manufacturer
NOA MEDICAL INDUSTRIES
Product Code
FNL
UDI-DI
00817492012745
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE LATCH IS A TWO STEP PROCESS WHERE THE BUTTON MUST BE PUSHED IN AND THE SIDE RAIL MOVED. THERE IS A POSSIBILITY THAT THEY DIDN'T GET THE SIDE RAIL PUSHED ALL THE WAY UP TO CAUSE THE BUTTON TO POP OUT AND THE RAIL TO LOCK INTO PLACE. A REPLACEMENT WAS SENT TO THE CUSTOMER. WE WERE WAITING FOR THE PART TO RETUN INCLUDING CONTACTING TWICE AND SENDING A RETURN LABEL. DEVICE WAS NOT RETURNED FOR INSPECTION SO A MALFUNCTION COULD NOT BE DETERMINED. IN HOUSE INVESTIGATION FOUND A SLIGHT TOLERANCE ISSUE ON THE FRAMES THAT ALLOWED THE LATCH ATTACHMENT PINS TO MOVE, CATCHING THE BUTTON ON THE COVER AND EITHER BREAKING IT OR HANGING UP ON THE COVER AND GETTING STUCK. BUTTON WAS ATTACHED TO THE LATCH WITH PLASTIC PINS THAT BROKE EASILY. THE RED BUTTON WAS REDESIGNED TO SHORTEN AND BE HELD IN PLACE WITH SCREWS INSTEAD OF PINS. CHANGED LATCH ATTACHMENT PINS TO A THREADED ROD AND PLATES WHICH SECURES THE LATCH TO PREVENT MOVEMENT. PIN LENGTH ON THE BACK OF THE LATCH WAS ALSO FOUND TO AFFECT PERFORMANCE. A CORRECTION IS PLANNED TO REPLACE FRAME ASSEMBLY (FRAME, LATCH, RED BUTTON).

Description of Event or Problem · 0

(B)(4), ON TUE, DEC 5, 2023 AT 9:41AM WE RECEIVED AN EMAIL ALLEGING THAT THE RED BUTTON ON THE SIDE RAIL THAT RELEASES THE RAIL TO DROP DOWN WAS STUCK INSIDE AND DOESN'T POP OUT AS IT SHOULD TO LOCK THE RAIL IN PLACE. THE RAIL ISN'T LOCKING INTO PLACE, WHICH ALLOWED HIM TO FALL OUT OF HIS BED DUE TO THE RAIL DROPPING FROM THE UPRIGHT POSITION. FORTUNATELY, HE IS OKAY BUT THE RAIL IS NOT. INCIDENT WAS REPORTED FROM A USER VIA A DISTRIBUTOR, NO OTHER INFORMATION WAS PROVIDED. WE SENT A REPLACEMENT WITH A RETURN LABEL SO THE DEVICE COULD RETURN FOR INVESTIGATION. WE REACHED OUT AGAIN LATER WHEN THE PART HAD NOT RETURNED OFFERING TO SEND ANOTHER RETURN LABEL BUT NEVER HEARD BACK FROM THE CUSTOMER. WE COULD NOT DETERMINE THAT A MALFUNCTION OCCURED. REPORT IS BEING FILED AFTER THE 30 DAY TIMELINE DUE TO CONFUSION OVER THE FILING OF MDRS WITH LITTLE INFORMATION. WE ARE REPORTING WITH THE INFORMATION REASONABLY KNOWN TO US OUT OF AN ABUNDANCE OF CAUTION. SIDE RAIL IS PART OF THE BED PACKAGE. THERE ARE NO PROBLEMS ASSOCIATED WITH THE BED ITSELF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
868974 NOA MEDICAL INDUSTRIES SIDE RAIL SET - FOOT BEHAVIORAL HEALTH BED (GRAY) FNL NOA MEDICAL INDUSTRIES 7540228LGRY 00817492012745

Patients

Seq Age Sex Outcome Treatment
1 NA Male Other