TECNIS SIMPLICITY
Report
- Report Number
- 3012236936-2025-000004
- Event Type
- Injury
- Date Received
- January 6, 2025
- Date of Event
- August 15, 2024
- Report Date
- March 26, 2025
- Manufacturer
- AMO PUERTO RICO MFG. INC.
- Product Code
- HQL
- UDI-DI
- 05050474731776
- PMA / PMN Number
- P980040
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
SECTION A3B: UNKNOWN/ ASKED BUT NOT AVAILABLE. SECTION H3: THE DEVICE WAS NOT RETURNED FOR EVALUATION. THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORD, COMPLAINT TRENDING, AND RISK DOCUMENTATION FOR THIS DEVICE WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ATTEMPTS HAVE BEEN MADE TO OBTAIN MISSING INFORMATION; HOWEVER, TO DATE, NO RESPONSE HAS BEEN RECEIVED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
ADDITIONAL INFORMATION: SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES. SECTION D9: RETURNED TO MANUFACTURER ON: 2/03/2025. SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: VISUAL INSPECTION REVEAL HALF A LENS PIECE WAS RECEIVED. THE LENS WAS CLEANED AND PRESENTED WITH NO FURTHER ISSUES. CONCLUSION: NO ISSUES THAT COULD CAUSE OR CONTRIBUTE TO THE COMPLAINT ISSUES REPORTED WERE IDENTIFIED DURING PRODUCT EVALUATION. THE ¿LENS CUT¿ OBSERVED DURING PRODUCT EVALUATION COULD NOT BE CONFIRMED TO BE RELATED TO THE MANUFACTURING OR DESIGN PROCESS. BASED ON THE COMPLAINT INVESTIGATION RESULTS, THE PRODUCT WAS RELEASED WITHIN SPECIFICATIONS. THE COMPLAINT ISSUE REPORTED COULD NOT BE CONFIRMED AND NO PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION COULD BE IDENTIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE PATIENT'S LEFT DUE TO BLURRY VISION, SHADOWS GHOST IMAGES, GLARES AND STARBURSTS. THE LENS WAS EXPLANTED IN SECONDARY PROCEDURE AND REPLACED WITH A NON JNJ LENS. THE PATIENT COULD NOT DRIVE AT NIGHT DUE TO VISUAL DISTURBANCES. THE BEST CORRECTED VISUAL ACUITY(BSCVA) PRE-OP WAS 20/50. THE BEST CORRECTED VISUAL ACUITY(BSCVA) POST-OP WAS 20/20. NO OTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 351600 | TECNIS SIMPLICITY | INTRAOCULAR LENS | HQL | AMO PUERTO RICO MFG. INC. | DIB00 | 05050474731776 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Female | Required Intervention| O |