PUMP 1886 780G OUS BLE PUMP MG/DL
Report
- Report Number
- 2032227-2025-103379
- Event Type
- Malfunction
- Date Received
- January 6, 2025
- Date of Event
- December 15, 2024
- Report Date
- April 8, 2025
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000521516
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. SUCCESSFULLY UTILIZED CREST AND THUS TO DOWNLOAD HISTORY FILES, TRACES AND COMLINK3 FILES. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM TRIGGERED PUMP ERROR 63. PUMP ERROR 63 ERROR DUE TO HARDWARE ERROR (FILE NUMBER: 49, LINE NUMBER: 19208 ESF#: 3926732 FOUND IN BOOTLOADER ERROR LOG. ALSO, PUMP ERROR 43 FOUND ON (B)(6) 2024 13:29:27, (B)(6) 2024 13:30:05, (B)(6) 2024 13:30:43, (B)(6) 2024 13:37:22. IN TRACES FILE HISTORY. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE ON BATTERY CONNECTOR, PCBA 1 / PCBA 2, FORCE SENSOR, MOTOR, VIBRATOR. FORCE SENSOR ZERO OFFSET WITHIN SPECIFICATION 22.8MV. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: UNIT HAD SCRATCHED CASE, CRACKED KEYPAD OVERLAY, STAINED KEYPAD OVERLAY, CRACKED BATTERY TUBE THREADS, CRACKED CASE CORNER OF BELT CLIP RAIL. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO PUMP ERROR 63 ALARM. PUMP ERROR 63 AND 43 ALARMS IN FILE HISTORY DUE TO MOISTURE DAMAGE ELECTRONIC ASSEMBLY. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED INSIDE THE UNITED STATES. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED PUMP ERROR 43(AN ERROR IN THE MOTOR WAS DETECTED BY READING AN UNEXPECTED VALUE FROM THE HALL SENSOR) AND CRITICAL PUMP ERROR (OPEN BOOK IMAGE). THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1886. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1886 WAS REQUESTED AND THE CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 962308 | PUMP 1886 780G OUS BLE PUMP MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1886 | NG3187150H | 000000763000521516 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |