FDA Adverse Event Injury Summary report: N

COBE CENTRYSYSTEM 3

MDR report key: 210701 · Received February 20, 1999

Report

Report Number
1713683-1999-00038
Event Type
Injury
Date Received
February 20, 1999
Report Date
January 20, 1999
Manufacturer
GAMBRO HEALTHCARE
Product Code
FII
Removal / Correction Number
NOT YET ASSIGNED
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

INFO REC'D THROUGH LITERATURE BY GAMBRO HEALTHCARE ("OUTBREAK OF BLOOD STREAM INFECTIONS IN A HEMODIALYSIS UNIT RELATED TO USE OF A DIALYSIS MACHINE WASTE PORT," ASN ABSTRACT, 31ST ANNUAL MEETING, OCTOBER 25-28, 1998, ATTACHED). THE MFR FILED 5 MDR'S WHEN THE INFECTIONS WERE FIRST LEARNED OF IN NOVEMBER. 1997 (REFERENCE 1714683-1997-00367 THROUGH 1713683-1997-00371), BUT THROUGH THIS LITERATURE LEARNED THAT 8 PTS WERE INVOLVED. ADDITIONAL MDR'S ARE NOW BEING FILED TO COVER THE REMAINING 3 PTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBE CENTRYSYSTEM 3 DIALYSIS CONTROL UNIT FII GAMBRO HEALTHCARE NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Hospitalization| R